Automated High-Performance Liquid Chromatographic Determination of Plasma Free Fatty Acids Using On-Line Derivatization With 9-bromomethylacridine Based on Micellar Phase-Transfer Catalysis

F A van der Horst, M H Post, J J Holthuis, U A Brinkman

J Chromatogr. 1990 Feb 2;500:443-52.

PMID: 2329146

Abstract:

The on-line use of micellar phase-transfer catalysis is described for the automated reversed-phase high-performance liquid chromatographic (RP-HPLC) determination of free fatty acids in plasma; minimum manual sample handling is involved. After diluting plasma ten-fold with the aqueous micellar system, which contains 25 mM of the non-ionic surfactant, Arkopal N-130 and 6 mM of the ion-pair agent tetrakis(decyl)ammonium bromide, the reaction of the fatty acids with the fluorophore 9-bromomethylacridine is complete within 5 min at 60 degrees C. Prior to RP-HPLC separation, interfering proteins are removed using an on-line filter and a column-switching unit. More than 100 samples can be injected onto a single pre-column. The detection limit is ca. 300 nM; the precision is better than 3% using an internal standard.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
AP14937429 Tetrakis(decyl)ammonium bromide Tetrakis(decyl)ammonium bromide 14937-42-9 Price
qrcode
Privacy Policy | Cookie Policy | Copyright © 2024 Alfa Chemistry. All rights reserved.