0

A New Potentiometric Electrode Incorporating Functionalized β-cyclodextrins for Diclofenac Determination

Joanna Lenik

Mater Sci Eng C Mater Biol Appl. 2014 Dec;45:109-16.

PMID: 25491808

Abstract:

This paper reports the preparation of diclofenac-selective membrane electrodes incorporating β-cyclodextrins: (2-hydroxypropyl)-β-cyclodextrin, heptakis(2,3,6-tri-O-methyl)-β-cyclodextrin, and heptakis(2,3,6-tri-O-benzoyl)-β-cyclodextrin. Several plasticized poly(vinyl chloride) membranes of different compositions were tested with the best electrode being the one incorporating heptakis(2,3,6-tri-O-benzoyl)-β-cyclodextrin with the membrane plasticized with 2-nitrophenyloctyl ether. The electrode is characterized by a near-Nernstian response slope of -60.0 mV decade(-1) over the linear range of 5.0×10(-5)-1.0×10(-2) mol L(-1) and a limit of detection of 1.4×10(-5) mol L(-1). The proposed electrode can easily discriminate diclofenac ions from several inorganic and organic interferents and some common drug excipients. The electrode has a response time of 10s and can be used within a pH range of 6.2-8.5 over 10 months without any considerable deterioration. The electrical properties of the membrane electrode were studied by impedance spectroscopy. The notable advantages of the diclofenac-selective electrode include its high sensitivity, selectivity, cost-effectiveness, and comfortable application in drug and urine analysis.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
LS7413161 Heptakis(2,3,6-tri-O-benzoyl)-β-cyclodextrin Heptakis(2,3,6-tri-O-benzoyl)-β-cyclodextrin Price
qrcode