Calibration Possibilities and Modifier Use in ETV ICP OES Determination of Trace and Minor Elements in Plant Materials

Albena Detcheva, Peter Barth, Juergen Hassler

Anal Bioanal Chem. 2009 Jul;394(5):1485-95.

PMID: 19468719

Abstract:

The possibilities for universal calibration based on multi-element aqueous standard solutions and graphite laboratory reference materials (graphite standards) for the electrothermal vaporization inductively coupled plasma optical emission spectrometric (ETV ICP OES) determination of Al, B, Ba, Cd, Co, Cr, Cu, Fe, Mn, Ni, P, Pb, S, Sr, Ti, V, and Zn in plant materials were investigated. A commercially available state-of-the-art ETV device was coupled with an Echelle ICP spectrometer equipped with a charge-injection-device (CID) camera for spectral detection. The transition area between transport tube and ETV graphite tube and the gas streams for inner gas, bypass gas, and modifier gas were optimized to achieve best transport efficiencies. The influence of four gaseous modifiers (CCl4, CHCl3, CCl2F2, and C3H8) added to the inner gas was studied. Five reference materials (RM P-Alfalfa, Lucerne; NIES CRM No.9 "Sargasso"; CTA-VTL-2 Virginia Tobacco Leaves; NIST SRM 1515 Apple Leaves; IAEA-V-10 Hay Powder) were used for method validation. If certified reference materials are not available, calibration against graphite standards or dried aqueous standard solutions is possible. Three carbonization procedures as sample pretreatment for the plant materials were investigated.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
AS2121447 Hay powder (elements) Hay powder (elements) Price