0

Comparative Study of Two Different Chromatographic Approaches for Quantitation of Hydrocortisone Acetate and Pramoxine Hydrochloride in Presence of Their Impurities

Fawzia Ibrahim, Asmaa Kamal El-Deen, Kuniyoshi Shimizu

J Food Drug Anal. 2018 Jul;26(3):1160-1170.

PMID: 29976408

Abstract:

In the present study, we compare the performance of two reversed-phase liquid chromatographic approaches using different eluents either conventional hydro-organic eluent or micellar one for simultaneous estimation of hydrocortisone acetate and pramoxine hydrochloride in presence of their degradants and process-related impurities; hydrocortisone and 4-butoxyphenol, respectively. For conventional reversed-phase liquid chromatography (RPLC), separation of the studied compounds was completed on an Inertsil ODS 3-C18 column (150 mm × 4.6 mm, 5 μm particle size) with a mobile phase consists of 50 mM phosphate buffer (pH 5.0): acetonitrile (50: 50, v/v). For micellar liquid chromatography (MLC), an Eclipse XDB-C8 column (150 mm × 4.6 mm, 5 μm particle size) was chosen for the separation with a green mobile phase consists of 0.15 M sodium dodecyl sulfate, 0.3% triethylamine and 10% n-butanol in 20 mM orthophosphoric acid (pH 5.0). Both methods were extended to analyze hydrocortisone acetate and pramoxine hydrochloride in their co-formulated cream. RPLC was superior to MLC with regard to sensitivity for the estimation of impurities. While, MLC represents an eco-friendly, less hazardous and biodegradable approach. Furthermore, the direct injection of the cream to the system without the need to laborious samples pretreatment, excessive amount of analysis time and/or use of large amount of toxic organic solvents is one of the outstanding advantages of MLC.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
AP122764801 Hydrocortisone phosphate triethylamine Hydrocortisone phosphate triethylamine 122764-80-1 Price
AP637581-A Pramoxine hydrochloride Pramoxine hydrochloride 637-58-1 Price
qrcode