0

Disposable Electrochemical Immunosensor for Cortisol Determination in Human Saliva

Satu Kämäräinen, Marianne Mäki, Tiina Tolonen, Giuseppe Palleschi, Vesa Virtanen, Laura Micheli, Adama Marie Sesay

Talanta. 2018 Oct 1;188:50-57.

PMID: 30029405

Abstract:

Cortisol is a steroidal hormone and an important stress marker. Free serum cortisol concentration has been identified to correlate well with free salivary cortisol. In this present work an electrochemical immunosensor was developed to determine cortisol concentration within the physiological concentration range found in human saliva. The immunosensor is based on a direct competitive enzyme linked immunoassay using a home-made cortisol-alkaline phosphatase (AP) conjugate synthesized in our laboratory with disposable graphite screen-printed electrodes (SPEs). 1-nalphtyl phosphate (1-NP) was used as an enzymatic substrate and a square wave voltammetry (SWV) for electrochemical detection. To study method suitability for use with saliva samples, calibration curves were performed both in buffer and saliva. In buffer standard samples showed a limit of detection (LOD) of 0.6 ng/ml and working range (WR) of 0.2-44.6 ng/ml with good reproducibility (RSD 10%). Saliva matrix effect was removed effectively with Salivette Cortisol collection device (polyethylene) and a calibration curve showed similar characteristics as in buffer with LOD 1.7 ng/ml and WR 0.5-55.1 ng/ml (RSD 8%) demonstrating the possibility to determine human salivary cortisol within the desired human physiological range. Spiked saliva samples were analyzed with the developed immunosensor presenting excellent 92-114% recovery. Comparison to liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) method showed strong 0.90 correlation between methods indicating good accuracy of the developed immunosensor.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
AS2121445 Human serum (cortisol spiked) Human serum (cortisol spiked) Price
qrcode