0

Fast Potentiometric Analysis of Lead in Aqueous Medium Under Competitive Conditions Using an Acridono-Crown Ether Neutral Ionophore

Ádám Golcs, Viola Horváth, Péter Huszthy, Tünde Tóth

Sensors (Basel). 2018 May 3;18(5):1407.

PMID: 29751498

Abstract:

Lead is a particularly toxic heavy metal that is present above acceptable levels in the water of many countries. This article describes a quick detection method of lead(II) ions using a polyvinyl chloride (PVC)-based ion-selective membrane electrode containing an acridono-crown ether ionophore by potentiometry. The electrochemical cell exhibits a Nernstian response for lead(II) ions between the concentration range of 10−4 to 10−2 M, and can be used in the pH range of 4⁻7. The applicability of this sensor was verified by measuring a multicomponent aqueous sample. Under the given conditions, this electrode is suitable for the selective quantitative analysis of lead(II) ions in the presence of many additional metal ions.

Chemicals Related in the Paper:

Catalog Number Product Name Structure CAS Number Price
AP90276587 Lead ionophore II Lead ionophore II 90276-58-7 Price