0

Incidence of Hypersensitivity Reactions During Hemodialysis

Raquel Esteras, Juan Martín-Navarro, Gabriel Ledesma, Raúl Fernández-Prado, Gilda Carreño, Melissa Cintra, Ignacio Cidraque, Ignacio Sanz, Blanca Tarragón, Simona Alexandru, Mónica Milla, Elena Astudillo, etc.

Kidney Blood Press Res. 2018;43(5):1472-1478.

PMID: 30235456

Abstract:

Background/aims:
A recent alert from Spanish health authorities warned of a higher incidence of reported hypersensitivity reactions to hemodialysis membranes with polysulfone, in the 2017 review of acute reactions to dialyzers found only published reports in the 21st century on polysulfone and its derivatives. The aim is to assess/evaluate the current incidence and characteristics of hypersensitivity reactions in hemodialysis patients.
Methods:
A retrospective multicentre study in 9 Spanish hospitals evaluated patients in whom a hypersensitivity reaction required a change in dialyzer membrane.
Results:
A total of 37 patients out of 1561 (2.37%) had hypersensitivity reactions and clinical, epidemiological and analytical data were available for 33 patients (2.11%). The membranes involved were polysulfone (n=23), polynephron (n=8), polyethersulfone (n=1) and polyacrylonitrile (n=1). This distribution reflected the frequency of use of membranes in the participating dialysis units. The reactions were described as type A in 18 cases and type B in 15 cases. There were no significant differences between the two types in clinical symptoms, the composition of the membrane involved, the method of sterilization, the season, or the time during the session in which they occurred. The most frequent symptom was dyspnea/breathlessness (64% of reactions). Eosinophilia was common (74%). 54% of the reactions occurred within the first 30 minutes of hemodialysis, 64% occurred during the first year of dialysis, and 54% required discontinuation of dialysis session. Cellulose triacetate was used as an alternative dialyzer in 78% of the cases.
Conclusion:
The incidence of hypersensitivity reactions was in the range found in reports from 20 years ago and is observed associated with synthetic membranes, not just polysulfones. Cellulose triacetate appears to be a good alternative for these patients.