Salt Tolerance Enhancement of Liquid Chromatography-Matrix-Assisted Laser Desorption/Ionization-Mass Spectrometry Using Matrix Additive Methylenediphosphonic Acid

Yuki Ohta, Shinichi Iwamoto, Shin-Ichirou Kawabata, Ritsuko Tanimura, Koichi Tanaka

Mass Spectrom (Tokyo). 2014;3(1):A0031.

PMID: 26819873

Abstract:

Mass spectrometry (MS) is a highly sensitive analytical technique that is often coupled with liquid chromatography (LC). However, some buffering salts used in LC (e.g., phosphate and tris(hydroxymethyl)aminomethane (Tris)) are incompatible with MS since they cause ion-source contamination and signal suppression. In this study, we examined salt tolerance of MALDI and applied a matrix additive methylenediphosphonic acid (MDPNA) to reduce salt-induced signal suppression. MDPNA significantly improved the salt tolerance of MALDI-MS. Using ammonium formate buffer at pH 5.0, the effective range of buffering salt concentration in MALDI-MS using MDPNA was estimated up to 250 mM. MDPNA reduced signal suppression caused by buffering salts at pH 4.0 to 8.0. We observed that MDPNA effectively worked over a wide range of buffer conditions. MDPNA was further applied to hydrophilic interaction chromatography (HILIC) and chromatofocusing-MALDI-MS. As a result, the analytes in the eluent containing high-concentration salts were detected with high sensitivity. Thus, our study provides simple and fast LC-MALDI-MS analysis technique not having strict limitation of buffering condition in LC by using matrix additive MDPNA.